Insulation Resistance Tester produces 10 kV test voltages to meet IEEE43-2000 Requirements
The new IEEE Standard 43-2000 entitled "IEEE recommended PRACTICE FOR testing Insulation Resistance of rotating Machines recommends using a 10 kV test voltage for motor windings rated greater than 12 kV." Additionally, the Polarization Index test is recommended. Designed to perform these tests in the insulation testing of : • Generators • Winding Motors • Traction Motors • Fan Motors
Valuable in industries such as: • Steel Manufacturing • Railroad and Mining • Electrical Power Generation • Rewind Shop • Heavy Industry
Features
Operates on internal rechargeable batteries or an AC supply
5 Voltage Ranges: 500, 1000, 2500, 5000, 10,000 Volts
Insulation Resistance Measurements to 500 GW 1 TW analog
Automatic Polarization Index Test
Data Storage and Downloading
Integral timer
IP54 Rating
Interference Rejection of 1 mA RMS per kV to a max. of 2 mA
Humidity Specification of 90% R.H. @ 40°C (104°F) maximum
SPECIFICATIONS
Dimensions
15 1/4 (385mm) x 10 1/2 (265mm) x 9 3/4 (250mm)
Weight
15 1/4 lb (7kg) approx.
Instrument Protection
IP54 with case closed
Safety
Meets EN61010-1; 2001 for operation at ≤ 2000 m above sea level
EMC
Meets EN61326-1:1998 for use in areas other than domestic
Test Voltages (d.c.)
500V, 1000V, 2500V, 5000V, 10000V
Test Voltage accuracy
± 5 % on 100Ω W loads and above
Insulation Resistance Ranges
Digital
10 k Ω to 50 G Ω @ 500 V; 10 k Ω to 100 G Ω @ 1000 V 10 k Ω to 250 G Ω @ 2500 V; 10 k Ω to 500 GΩ @ 5000 V 10 k Ω to 500 G Ω @ 10000 V
Analog
100 k Ω to 1T Ω @ all voltages
Accuracy (at 23° C)
± 5 % on 100 M Ω load
Basic Accuracy
± 5 % of reading 10 M Ω to 100GΩ all voltages
Current Range
0.01 nA to 999 μA
Accuracy (23 deg C)
± 5 % ± 1nA on 100 MΩ load
Basic Accuracy
± 5 % of reading 10M Ω 100 GΩ all voltages
Guard
2% error guarding 500 k Ω leakage with 100MΩ load
Short Circuit Current
5mA nominal
Voltmeter Range
50 to 1000V dc or ac
Accuracy (at 23° C)
± 2 % ± 1V
Interference Rejection
1mA RMS per kV to a maximum of 2mA
Capacitor Discharge Time
<2 s per μF to discharge from 10kV to 50V
Timer
0:00 to 99:59 minutes. Can be set to stop test after time limit or run indefinitely
Instrument Display
3 digit LCD Analog/Digital
Power Supply
Battery or Line Powered; Rechargeable lead-acid batteries (12,V 4Ah), or 100V to 240VAC, 50/60 Hz supply
Battery Life
Typically 4 hrs. continuous testing @ 10kV; Full charge in approximately 12 hrs
Temperature Range
Operating: -20° C to +50°C (-4° F to + 122° F); Storage: -25° C to + 65°C (-12° F to + 149° F)
Temperature Coefficient
±0.1% per °C over the test voltage for the temperature range 0° C to +30°C (+ 9.06% per °F)
Humidity
90% RH @ 40° C (104° F) max.
CAT NO.
MODEL
DEscRIPTION
AVAIL
PRICE
BD21001
MEG10-01
10 kV Insulation Tester; includes test leads, RS232 cable, User Guide, Download Manager CD-ROM, Quick Start Guide, Power Cord
Yes
10,692,000